Several Maintenance Methods Of Vacuum System Of Freeze Dryer

Apr 11, 2022

Leave a message

1. How to adjust the expansion valve


The clockwise direction is to reduce the flow of the expansion valve; the counterclockwise direction is to increase the flow of the expansion valve; The amount of adjustment each time should not be too large, and the time interval between two times should not be less than 15 minutes, so as to observe the running trend of the system. Once the expansion valve is set, do not move it unless the operating conditions change greatly.


2. How to collect liquid


First close the liquid outlet valve of the water condenser, start the compressor, open the plate cooling solenoid valve and condensing solenoid valve for 10s, run the compressor for 5-8min, check the medium pressure of the compressor, and stop the compressor when the medium pressure is about 0kgf/cm At the same time, the inlet valve of the water condenser is closed, and the freon is collected into the water condenser.


FDF35


3. Replace the cleaning filter


It is recommended to replace the filter element of the drying filter and the filter element of the return air filter after the refrigeration system has been running for more than 1 year. It is best to replace it once a year. If it cannot be replaced, it should be removed and cleaned with gasoline.


4. How to discharge the air from the refrigeration system


Close the liquid outlet valve of the water condenser, start the compressor, open the plate cooling valve and the condensing valve after 10s, and turn off the compressor after 5min of continuous operation. For the inlet valve of the water condenser, unscrew the nut next to it and turn the valve stem clockwise to discharge air.